World Precision Instruments Biological Signals Amplifier Manuel d'instructions

Naviguer en ligne ou télécharger Manuel d'instructions pour Amplificateurs audio World Precision Instruments Biological Signals Amplifier. World Precision Instruments Biological Signals Amplifier Instruction manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 114
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
INSTRUCTION MANUAL
050707
WARNING: Do not connect the
Lab-Trax hardware to your
computer until AFTER the
software installation is complete.
Data-Trax2
Data recording and analysis software & Lab-Trax™ data acquisition systems
World Precision Instruments
www.wpiinc.com
Vue de la page 0
1 2 3 4 5 6 ... 113 114

Résumé du contenu

Page 1 - Data-Trax2

INSTRUCTION MANUAL050707WARNING: Do not connect the Lab-Trax hardware to your computer until AFTER the software installation is complete.Data-Trax2™Da

Page 2

. . Introduction 4

Page 3

. . 6 Input and Output 94 Figure 6-18: Channels page of the Preferences dialog window set for testing an Output Sequence.Stimulator Page Figure 6-19:

Page 4

6 Input and Output 95On the Stimulator page, select Pulses as the stimulating mode and set the parameters.• Toolbar StepsFreq (Hz) = 1; Amp (V) = 0.0

Page 5 - Controls 35

. . 6 Input and Output 965 Return to the Main window; select the CAPRecruit sequence from Sequences control on the toolbar. 6 Click the Record button

Page 6 - Export 104

6 Input and Output 97 Figure 6-21: Output StatesSequences Digital states can be grouped together in Sequences. A Sequence allows timed execution of d

Page 7 - Introduction

. . 6 Input and Output 98Firing a Sequence The Sequence Control on the Main toolbar is a menu containing sequences of analog or digital outputs defin

Page 8 - Technical Support

6 Input and Output 99EventsAs DataTrax2 acquires data, it is aware of the value of each data point as it happens. It should then be possible to "

Page 9 - • Note your computer model

. . 6 Input and Output 100Timed Events Timed events are triggered after a certain amount of time has passed. Timed events can be repeated, set N = 0

Page 10 - Introduction 4

7 Using Settings 101Chapter 7: Using SettingsThe Settings MenuDataTrax2 offers users several choices for recording and displaying data. When certain

Page 11 - Chapter 1: Quick Start

. . 7 Using Settings 102Creating a New Settings FileThe first step in creating a settings group is to create a settings file. Configure the DataTrax2

Page 12 - Dialog Pages

7 Using Settings 103Categories Settings files can be organized into categories within a group. To create a new category in the settings group, click

Page 13 - 1 Quick Start 7

1 Quick Start 5Chapter 1: Quick StartStart here if you are already familiar with data recording concepts. This chapter contains all you need to know

Page 14 - 1 Quick Start 8

. . 8 Export 104Chapter 8: ExportPrintingOnce you have recorded and analyzed your data, you’ll probably want to create a report or other presentation

Page 15 - 1 Quick Start 9

8 Export 105 Figure 8-1: Text and image as displayed in the Journal window.Exporting DataData can be exported from DataTrax2 as either numbers or pic

Page 16 - Chapter 2: Acquisition

. . 8 Export 106If data is exported from the Analysis or ScopeView windows, then only the data displayed in the window is exported.Data from the Anal

Page 17 - 2 Acquisition 11

AAdd Data to Journal 37aliasing 10Amplitude 23Analysis 7Area 73Power 51Analysis window 8, 37, 39, 104ASCII 8Auto Find 37Auto Mark 37Autocorrelation 58

Page 18 - Signal Conditioning

Multiple Sweeps 17Multipoint 56NNational Instrumentshardware 27Navigating Data 7Ooscilloscope 16PPaste 8Peak Analysis 66Period 7, 45png 8Power 7, 9Pre

Page 19 - 2 Acquisition 13

. . 1 Quick Start 6• Clicking the Double Display Time button (little mountain) doubles the screen time. Therefore, a 10-second, full-screen display b

Page 20 - Chart Mode

1 Quick Start 7Channel Mode Clicking on Channel Mode in Raw Data displays the input option available for a given channel. On a computed channel, cli

Page 21 - 2 Acquisition 15

. . 1 Quick Start 8a single vertical line in Figure 1-1 on page 5). 4 The text attached to a mark can be seen at the bottom of the screen, just above

Page 22 - Scope Mode

1 Quick Start 9On a given selection of data displayed in the Analysis window, the program can calculate and report the selected parameters. These are

Page 23 - 2 Acquisition 17

. . 2 Acquisition 10Chapter 2: AcquisitionThe Basics Time, Temporal Resolution, Sample Speed, and Aliasing Temporal resolution in digitally recorded

Page 24 - Saving Your Data

2 Acquisition 11connected by a line, a sine wave with a frequency of 0.1Hz is displayed on the screen. Again, the wave displayed on the screen is not

Page 25 - 2 Acquisition 19

. . 2 Acquisition 122 Open the Channels page in the Preferences dialog window and set the sampling speed to 1,000 samples per second. Record 10 secon

Page 26 - Chapter 3: Display

2 Acquisition 13For example, consider an ECG, whose total peak to peak amplitude is only 2 mV. The number of steps that can be resolved with an 16 bi

Page 28 - Behavior of

. . 2 Acquisition 14mains. The end result of using the 50Hz Low Pass filters is the creation of a quieter, more readable ECG. In general, the applica

Page 29 - Display Time

2 Acquisition 15The dialog boxes (Figure 2-2 on page 15) show the options for starting the recording process for both groups of hardware. Figure 2-2:

Page 30 - 3 Display 24

. . 2 Acquisition 16a small piece of data prior to the trigger. To enable pretriggering, check the box next to the label Pretrigger (in the Start Mod

Page 31 - 3 Display 25

2 Acquisition 17not show the event that was trying to be captured. So the acquisition of data at the right moment requires a proper trigger. In the S

Page 32 - 3 Display 26

. . 2 Acquisition 18Sweep Length When operating in Scope mode, DataTrax2 takes “snapshots” of data. Each “snapshot” is called a sweep and has a pre-d

Page 33 - Converting

2 Acquisition 19

Page 34 - 3 Display 28

. . 3 Display 20Chapter 3: DisplayIntroductionFrom a functional standpoint, the DataTrax2 program is really two programs. It is primarily a data acqu

Page 35 - 3 Display 29

3 Display 21Each channel has its own set of controls located in a Channel Bar immediately above each channel’s data window (Figure 3-2 on page 21). Y

Page 36 - Making Marks

. . 3 Display 22for the position of the cursor is reported in the Clock Area in the upper right hand corner of the Main window. Single Cursor mode is

Page 37 - Off-Line

3 Display 23Main Window Display ConsiderationsRecorded data has two important dimensions: length and height. The length is the Time axis and the heig

Page 38 - Comments

You are free:to Share — to copy, distribute and transmit the workto Remix — to adapt the workUnder the following conditions:Attribution. You must attr

Page 39 - Exporting Marks

. . 3 Display 24 Figure 3-5: ECG viewed with a long screen time (top) and a short screen time (bottom).In the second example (Figure 3-6 on page 24),

Page 40 - Figure 3-15: Online XY

3 Display 25Managing Amplitude DisplayThe vertical display of your signal is managed in a number of ways. In each channel toolbar (Figure 3-7 on page

Page 41 - Chapter 4: Controls

. . 3 Display 26Rate function could be used on Channels 9-16 to calculate the heart rate of each animal from its recorded blood pressure on Channels

Page 42 - Figure 4-3: The View menu

3 Display 27Note: Any changes to the default view, will also be made to the preferences and vice-versa.Channel Sizing Figure 3-8: The height of the d

Page 43 - 4 Controls 37

. . 3 Display 28 Figure 3-9: Data file of weights hung from a transducer and the units conversion dialog window that sets volts to grams.Once recordi

Page 44 - Figure 4-6: The Help menu

3 Display 29Cursors 1 and 2. Enter the corresponding values in real units into the two value boxes on the right. In the Name area, enter the name of

Page 45 - Toolbars

. . 3 Display 30Units can also be set from the Channel page of the Preferences dialog window. Each channel can be set with the conversion factors pro

Page 46 - Pulse Mode

3 Display 31Preset Marks If you know the marks that you will be adding to the record prior to recording, you can store the mark text as preset marks

Page 47 - Step Mode

. . 3 Display 32Navigating By MarksMarks that are placed on the recording can serve as “sign posts,” indicating where important sections of data are

Page 48 - Dialog Windows

3 Display 333 Wait one minute, type “Test 2" and strike the Return or Enter key.4 Click Stop. Scroll through the data using the scroll bar on th

Page 50 - Chapter 5: Analysis

. . 3 Display 34 Figure 3-15: Online XYDisplay Considerations in Other Windows Data is only recorded in the Main window. However, other windows can r

Page 51 - 5 Analysis 45

4 Controls 35Chapter 4: ControlsMenusFile The DataTrax2 software supports all of the elements found in a standard File menu: Figure 4-1: The DataTra

Page 52 - Transforms

. . 4 Controls 36Edit The DataTrax2 software supports elements found in a standard Edit menu: Figure 4-2: The Edit menu.Undo: Un-does the last comma

Page 53 - 5 Analysis 47

4 Controls 37OnlineXY: Displays realtime XY plot to the right of the Main window channels.Main Window: Brings the Main window to the foreground. Also

Page 54 - 5 Analysis 48

. . 4 Controls 38DownSample: Used to downsample the data. This reduces the information in the file and should be used with caution. Warning: This can

Page 55 - Channel

4 Controls 39Check for Updates: Checks the DataTrax2.com website for the latest version of the software. You need to be connected to the internet.Too

Page 56 - Channel Math

. . 4 Controls 40Clicking on the down arrow next to the view name, brings up the View control menu. You can switch the current view from this menu, c

Page 57 - 5 Analysis 51

4 Controls 41•Apply All: Applies the changes and fires all stimulators. As with Apply, on hardware that supports it, the change will happen at the ne

Page 58 - Functions

. . 4 Controls 42Apply: Applies any changes to the stimulus protocol made through the Stimulator Toolbar. Clicking Apply while recording will send a

Page 59 - 5 Analysis 53

4 Controls 43Units: Used to convert two raw data values to two corresponding calculated values, using a linear scale. For more information refer to t

Page 60 - 5 Analysis 54

DataTrax2 Manual i Introduction 1Welcome 1How to Use This User’s Guide 1System Requirements 1Installation 1Registration and Updates 2Comments and Sugg

Page 61 - Calculation

. . 5 Analysis 44Chapter 5: AnalysisOverviewToday’s analog to digital converters make acquiring data very easy. The challenge to all users is to extr

Page 62 - Calibration

5 Analysis 45Note: Deleting the raw data channel will cause the function channel to have invalid data.Periodic Periodic functions operate on cyclic d

Page 63 - Template

. . 5 Analysis 46Exercise To see Threshold and Tolerance in use, record some pulse data using the Tutorial setting from one of the Settings groups. A

Page 64 - Correlation

5 Analysis 47Negative: Only the negative data points are included in the calculation.Data points with values above zero make the Integral larger, tho

Page 65 - Analysis

. . 5 Analysis 48 Figure 5-3: Integral with ZeroThe Reset control in the Integral Setup dialog window determines when the Integral or running count w

Page 66 - 5 Analysis 60

5 Analysis 49Also enter this X value in the text box below the graph. Clicking the Calibrate Difference between Cursors button will calculate the req

Page 67 - 5 Analysis 61

. . 5 Analysis 50unclosed parentheses or division by zero. Since division by zero is a possible occurrence even in legitimate expressions, but it can

Page 68 - 5 Analysis 62

5 Analysis 51Notice that the filter is subject to the Nyquist limitation of frequency. The maximum frequency in the Filter Setup dialog is exactly ha

Page 69

. . 5 Analysis 52known as a Power Number. In addition to specifying the range of frequencies transformed, the frequency resolution can be selected. T

Page 70 - 5 Analysis 64

5 Analysis 53In this function, a preset frequency band is available for each component of the EEG: Alpha, Beta, Theta, Delta, Betalow, Beta Mid and B

Page 71 - Blood Pressure

ii DataTrax2 ManualChapter 5: Analysis 44Overview 44Analyses of the First Kind: Wave-to-Wave Transforms 44Analyses of the Second Kind: Digested Data 5

Page 72 - Peak Analysis

. . 5 Analysis 54Digital Input The Digital Input function treats the data on the selected channel as a binary number.The functions in the Digital Inp

Page 73 - Copy, Export

5 Analysis 55CrossTime Channel CalculationInitial Delay: The amount of time from the start of recording until the function is initiated.Selection Dur

Page 74 - Components

. . 5 Analysis 56MultiPoint CalibrationMultiPoint Calibration can be used to perform a linear and non-linear calibration of sensors, transducers, amp

Page 75

5 Analysis 57Template MatchTemplate Matching is a technique in digital signal processing for finding small parts of a signal which match a template.

Page 76 - 5 Analysis 70

. . 5 Analysis 58Auto CorrelationCorrelation is the mutual relationship between two or more random variables (Ali). AutoCorre-lation is the correlati

Page 77 - 5 Analysis 71

5 Analysis 59Analyses of the Second Kind: Digested DataThe functions in this kind of analysis take a group of data points selected by the user and &q

Page 78 - FFT Functions

. . 5 Analysis 60 Figure 5-7: The DataTrax2 Analysis window.Analysis Window ComponentsFunctions List The functions selected in the Add Functions list

Page 79 - Figure 5-9: FFT channel menu

5 Analysis 61Channel Menu in the Analysis WindowClicking on the Channel Menu button or right-clicking in the display area of the Analysis window brin

Page 80 - Find Dialog

. . 5 Analysis 62V2-V1: Value2-Value1, i.e. the difference in value between the cursorsT2-T1: Time2-Time1, i.e. the difference in time between the cu

Page 81 - Auto Find

5 Analysis 63SpirometryTV: Tidal VolumeIRV: Inspiratory Reserve VolumeERV: Expiratory Reserve VolumeFEV1: Forced Expiratory Volume in one second. How

Page 82 - Specific

Introduction 1 IntroductionWelcomeThank you for choosing WPI DataTrax2 Data Recording and Analysis Software. Developed over a 5-year period, DataTra

Page 83 - 5 Analysis 77

. . 5 Analysis 64ECGTime at R: Time at the peak of the R-wave.PR Interval: The time from the beginning of the P wave to the peak of the R wave.P Dura

Page 84

5 Analysis 65Blood PressureLVEDP pressure: Left Ventricular End Diastolic PressureTime at LVEDP: Time at LVEDPMax Pressure: Maximum of the pressure,

Page 85 - 3rd, or 4th Kinds of Analyses

. . 5 Analysis 66Cycle Duration: Time between LVEDPs Relaxation Index: (Minimum Slope) / (Pressure at Minimum Slope)Tau (Mirsky): Time for pressure t

Page 86 - 5 Analysis 80

5 Analysis 67Trailing Slope: Average slope of the signal as it travels from the upper threshold to the lower thresholdMax Slope: Maximum slopeMin Slo

Page 87 - Chapter 6: Input and Output

. . 5 Analysis 68XY View Window ComponentsSelecting Channels To plot an XY graph, select the X-axis channel from the drop-down box in the top most pl

Page 88 - Protocols

5 Analysis 69Table Functions These functions are available in the XY window.General Y1: Value of Y-Axis Channel at Cursor 1.Y2: Value of Y-Axis Chann

Page 89 - Continuous Pulses

. . 5 Analysis 70Channel Menu in the XY WindowClicking on the Channel Menu button or right-clicking in the display area brings up the Channel menu. A

Page 90 - Protocol

5 Analysis 71the maximum frequency that could be recorded reliably would be 500Hz. If a recording was made at 100 samples per second, the maximum fre

Page 91 - 6 Input and Output 85

. . 5 Analysis 72 Figure 5-8:FFT WindowDisplay Channel Located in the upper left of the channel display area, this drop-down box allows the user to

Page 92 - Pulse Trains

5 Analysis 73Table Functions These functions are available in the FFT window.General Power1: Power at Cursor 1Power2: Power at Cursor 2Freq1: Frequen

Page 93 - 6 Input and Output 87

. . Introduction 21 Insert the DataTrax2 installation CD. 2 A welcome screen will appear offering to install Software.3 Follow the instructions pres

Page 94 - Voltage Steps

. . 5 Analysis 74Analyses of the Fourth Kind: Detected Data ConditionsDataTrax2 software can be used to identify Regions Of Interest (ROI) in recorde

Page 95

5 Analysis 75• Copying values and their headings to the Journal via the Add Title to Jrnl and Add Data to Jrnl functions in the Right-Click menu of t

Page 96 - Triangle

. . 5 Analysis 76After the number of repetitions specified or the specified end of the Auto Find routine is reached, the Auto Find routine can be sav

Page 97

5 Analysis 77 Figure 5-11: ECG AutoMark Dialog Figure 5-12: ECG AutoMark Example

Page 98 - Triangle Mode

. . 5 Analysis 78Blood Pressure Figure 5-13:Blood Pressure Automark Dialog Figure 5-14: Blood Pressure Automark Example

Page 99 - 6 Input and Output 93

5 Analysis 79Peak AnalysisAnalysis of the Fifth Kind: Things Not Possible with 1st, 2nd, 3rd, or 4th Kinds of AnalysesNot Yet Implemented in DataTrax

Page 100 - Stimulator Page

. . 5 Analysis 80For example, we have created a script in Python, which operates on the R-R intervals of the ECG, which are detected by an Auto Mark

Page 101 - Sequence Page

6 Input and Output 81Chapter 6: Input and Output StimulatorThe Stimulator page in the Preferences dialog box controls the Digital-to-Analog Converter

Page 102 - Outputs States

. . 6 Input and Output 82Pulse Protocols Figure 6-1: The parameter boxes on the Stimulator page of the Preferences dialog window.In order to understa

Page 103 - Figure 6-21: Output States

6 Input and Output 83Holding Potential: This is a voltage between ±10V that can be programmed to shift the resting voltage of an excitable tissue (l

Page 104 - Digital Input

Introduction 3• Note your computer model.• Note your operating system version.• Note the amount of RAM (Random Access Memory) in your computer.• Not

Page 105 - 6 Input and Output 99

. . 6 Input and Output 84Holding Potential 0 VThese settings will create a protocol that delivers continuous square waves with 1V amplitude and 100Hz

Page 106 - 6 Input and Output 100

6 Input and Output 85 Figure 6-4: The parameter boxes on the Stimulator page for Train mode.The terms used on the Stimulator page of the Preferences

Page 107 - Chapter 7: Using Settings

. . 6 Input and Output 86Holding Potential: This is a voltage between ±10V that can be programmed to shift the resting voltage of an excitable tissue

Page 108 - 7 Using Settings 102

6 Input and Output 87 Figure 6-5: Burst of pulses separated by Inter Train Durations.Constant Voltage ProtocolSelecting the Constant (voltage) option

Page 109 - 7 Using Settings 103

. . 6 Input and Output 88If the user knows the voltage increment needed at each step, the equation can be transposed to solve for the Number of Steps

Page 110 - Chapter 8: Export

6 Input and Output 896 To create a step protocol where the voltage returns to a baseline value between steps, alter the Time Off to a number greater

Page 111 - Exporting Data

. . 6 Input and Output 90Delay Amplitude: The amplitude during the delay period. Only available on the WPI 3xxStart Amplitude: The starting amplitude

Page 112 - 8 Export 106

6 Input and Output 91Holding potential: The amplitude during the interprotocol duration. Should be set to Start Amplitude if not being usedThe Stimul

Page 113

. . 6 Input and Output 92Step Mode Figure 6-14: The Stimulator Step Control Panel.A1: Starting AmplitudeA2: Stopping Amplitude.#steps: Number of Step

Page 114

6 Input and Output 93Rise time: The time taken to go from the Start Amplitude to the Stop Amplitude# Triangles: Number of TrianglesIR: Time between T

Commentaires sur ces manuels

Pas de commentaire